LABOTHERM

Laminar flow hood Two 30

PriceOn quote● Supplier reply within 48 h

Request a quote for this product only, or add it to your selection to group several products and enter your details just once. Free and without obligation.

Why choose our Marketplace ?

  • Simplify your purchases
  • Over 20 years of expertise
  • Certified suppliers
  • Fast and free service
Logo LABOTHERM LABOTHERMManufacture, complete laboratory installations, plastics boilermaking.

Description

Laminar flow hood Two 30

Advantages
product protection

transparent side panels

can be operated at either slight positive or negative pressure

compliant with standard NF X 44 101

Applications
Microbiology

Sterility testing

Cell culture

Sterile handling

PCR, DNA work

Principle
Protection of the work volume to ISO 5 class (NF EN 14644-1).

This hood protects the product from the operator.

Compact, economical and easy to install, it offers two operating modes:

Positive pressure

Part of the air exits through the front opening

Negative pressure

A portion of the filtered air is extracted through a pre-filter providing a slight protective barrier at the front

Construction

Epoxy-painted steel casing

Perforated work surface in brushed stainless steel

Laminated safety glass side panels

Electronically controlled fan

HEPA filters of type H14 (efficiency > 99,999% at 0,3 ?)

Controls

The control panel, accessible from the front, includes:

- on/off switch

- lighting switch

- UV switch (option)

- power socket switch (option)

- flow reduction switch

- anemometer (option)

- fuses

Other standard equipment

Lighting

by fluorescent tubes located above the work volume, outside the sterile zone

Work surface

made of perforated stainless steel in 1 piece